lcms
Maintenance
To reinstall the electrospray nebulizer1
To reinstall the electrospray nebulizer
When required
Tools required
Parts requiredAs necessary NoneNone
1Insert the nebulizer part way into the spray chamber.
CAUTIONDo not hit the tip of the needle as you insert the nebulizer. The tip of the needle is easily damaged.
2Reconnect the nebulizing gas tubing to the nebulizer.
3Finish inserting the nebulizer into the spray chamber.
4Turn the nebulizer clockwise and lock it in place.
5Reconnect the LC tubing the nebulizer.
CAUTIONDo not overtighten the LC fitting. Overtightening the fitting can crush the tubing, creating a restriction.
6Close the nebulizer cover.