lcms
1Maintenance
To clean the electrospray spray chamber weekly
To clean the electrospray spray chamber weekly
When required
Tools requiredWeekly or whenever symptoms indicate that there is contamination in the spray chamber and normal daily cleaning does not correct the problem Abrasive paper, 8000 grit (p/n 8660-0852)
Cloths, clean, lint-free (p/n 05980-60051)
Cotton swabs (p/n 5080-5400)
Gloves, clean
Isopropanol, reagent grade or better
Mobile phase, current
Wash bottle, clean
Water, reagent-grade or better
Parts requiredNone
Recent residue should be soluble in the mobile phase. If you are not sure what mobile
phase was used recently, a mixture of 50% isopropanol and 50% water works well as a
general cleaning solution. NOTE
1Turn off the spray chamber.
WARNINGThe electrospray spray chamber operates at high temperatures. Allow sufficient
time to cool down.
2Remove the electrospray nebulizer.
3Open the spray chamber and remove it from the LC/MS.
4Fill the spray chamber with clean mobile phase, or with a mixture of
isopropanol and water.
百度搜索“77cn”或“免费范文网”即可找到本站免费阅读全部范文。收藏本站方便下次阅读,免费范文网,提供经典小说教育文库Agilent三重四极杆LCMS维护手册(10)在线全文阅读。
相关推荐: