lcms
Maintenance
To adjust the electrospray nebulizer needle1
To adjust the electrospray nebulizer needle
When requiredAfter replacing the electrospray nebulizer needle or if symptoms indicate the
needle is not correctly adjusted
Tools required Adjustment fixture (p/n G1946-20215)
Gloves, clean
Magnifier (p/n G1946-80049)
Wrench 3-mm, open-end (p/n 8710-2699)
Wrench ¼-inch x 5/16-inch, open-end (p/n 8710-0510)
Parts requiredNone
1Install the nebulizer in the adjustment fixture. Figure 9
2Loosen the needle holder locknut.
3Position the magnifier so you can see the tip of the nebulizer.
4Adjust the needle holder until the needle is even with the tip of the
nebulizer.