lcms
1Maintenance
To adjust the electrospray nebulizer needleFigure 10
Adjusting the needle holder
5Tighten the locknut. Make sure this does not change the position of the
needle.
6Remove the nebulizer from the adjustment fixture and reinstall it in the
electrospray spray chamber.
7Be very careful not to hit the tip of the nebulizer against anything. Any
damage will have a large, negative effect on system performance.
NOTERecord this procedure in the Maintenance Logbook.