1Maintenance

To replace the electrospray nebulizer needle

To replace the electrospray nebulizer needle

When required

Tools required

Parts required

When the needle is plugged. Common symptoms of a plugged needle are increased LC back pressure or off-axis spraying or dripping from the nebulizer. Adjustment fixture (p/n G1946-20215) Gloves, clean Pliers, long nose (p/n 7810-0004) Wrench 3-mm, open-end (p/n 8710-2699) Wrench ¼-inch x 5/16-inch, open-end (p/n 8710-0510) Nebulizer needle kit, Electrospray (includes needle, ferrule, and needle holder, p/n G2427A) Figure 6

Electrospray nebulizer needle in needle holder1Install the nebulizer in the adjustment fixture. 2Loosen the locknut next to the zero-dead-volume (ZDV) union. 3Remove the union from the nebulizer. 4Loosen the locknut of the needle holder. 5Unscrew the needle holder and pull it out of the nebulizer. 6Slide the non-tapered end of the needle through the new needle holder from the narrower side.